ISO/IEC 17025 guide
ISO 17025 scope of accreditation: fixed and flexible scopes, and how to write yours
Clause by clause: ISO/IEC 17025:2017 requirements: every clause explained →
What a scope of accreditation lists, how to word each line of the schedule, when a fixed or a flexible scope suits your laboratory, what can stay off the application and how the scope sets assessment effort.
Key points
- An ISO 17025 scope of accreditation is the public schedule listing the tests or calibrations, methods, ranges and sites the accreditation body has assessed. Anything not on it is not accredited.
- You do not remove requirements of the standard from your scope. You choose which laboratory activities to put on the schedule, and every requirement applies in full to those activities.
- A fixed scope names each method and version; a flexible scope lets the laboratory add or update methods inside agreed boundaries, under the body's own rules and a procedure the assessors have tested.
- The scope drives the number of technical assessors, the witnessing and the assessment days, so write it for the results customers and regulators actually rely on.
What is an ISO 17025 scope of accreditation?
A laboratory is never accredited in general. The accreditation body grants accreditation for a defined list of laboratory activities, publishes that list alongside the certificate, and assesses the laboratory against it at every surveillance visit and reassessment.
The requirement behind it sits in clause 5, Structural requirements: the laboratory defines and documents the range of laboratory activities for which it claims conformity with ISO/IEC 17025. The body turns the part you apply for, once assessed, into the published scope.
This guide answers one question: what your scope should contain and how to word it. Whether you need accreditation at all is covered in the decision guide for laboratories, and the order of events from application to decision is in the accreditation process guide.
What goes into each line of the schedule
Accreditation bodies set their own schedule layouts, and a calibration schedule looks different from a testing one, but the same building blocks appear.
| Element | What to state | What goes wrong |
|---|---|---|
| Discipline or field | The area of work the body uses to group activities and allocate technical assessors, such as chemical, microbiological, mechanical or a calibration quantity | Mixing disciplines on one line, so no single assessor can cover it |
| Item, material or instrument | What is tested or calibrated: the product, matrix, sample type or instrument class | "All materials" or "various samples", which no validation file supports |
| Test, property or quantity | What is determined or measured, in the terms your reports use | Naming a technique instead of a result |
| Method | The standard method and its version, or the in-house method reference | Citing a standard your procedure has quietly departed from |
| Range | The range over which the method has been verified or validated | A range copied from the standard rather than from your own data |
| Measurement capability (calibration) | The capability you claim, expressed as the body requires | A figure your uncertainty budget cannot support |
Each line needs evidence behind it: a verification or validation file, an uncertainty evaluation, authorised staff, traceable equipment and, where a scheme exists, proficiency testing. The detail of that evidence belongs to the guides on method verification and validation, measurement uncertainty and proficiency testing.
Fixed and flexible scopes
A fixed scope lists each method exactly, usually with its version. Any new method, new matrix, wider range or revised standard is an extension: the laboratory applies, the body assesses, and the schedule is reissued before the new activity may be reported as accredited.
A flexible scope describes a boundary rather than a list, for example a technique applied to a defined group of materials, and permits the laboratory to bring new or revised methods inside that boundary itself. The trade is clear: the laboratory gains speed, and in exchange it must run a documented, validated procedure for adding methods, keep a current internal list of what is covered, and show the assessors that each addition was handled correctly.
Whether flexibility is available, in what forms and with what limits, is decided by each accreditation body, not by ISO/IEC 17025. UKAS (in its GEN 4 provisions) and ANAB (in PR 2311, Flexible Scopes) are two bodies that publish rules on it. Read your body's document before you design your scope around it.
| Question | Fixed scope | Flexible scope |
|---|---|---|
| What the schedule shows | Each method, version and range | A defined boundary, with the laboratory keeping the detailed list |
| Adding a method or new version | Extension assessed by the body before use as accredited | Added by the laboratory under its approved procedure, reviewed at the next visit |
| Suits | Stable methods, first accreditations, small laboratories | Laboratories with a strong validation track record and frequent method change |
| Main risk | Waiting for extensions when standards are revised | An addition that was not properly validated, found at surveillance |
For a first accreditation most laboratories are better served by a fixed scope.
What can be left out, and what cannot
ISO/IEC 17025 does not work like a management system certificate where some requirements are declared not applicable. The standard applies to all organisations performing laboratory activities, whatever their size. What you choose is which activities appear on the schedule; for those activities, clauses 4 to 8 apply in full. The one structural choice in the standard itself is in clause 8, where the laboratory runs its management system under Option A or Option B (a system meeting ISO 9001).
- You may leave out activities. Tests you perform rarely, methods still being validated or work nobody asks to be accredited can stay off the application. They remain unaccredited, and reports for them must not suggest otherwise.
- You may not leave out a requirement for an activity you list. If sampling is part of the listed activity, the sampling requirements (7.3) apply. If you issue statements of conformity, the decision rule agreed in contract review (7.1) and the reporting rules (7.8.6) apply.
- Work you send out is not on your scope. Tests carried out for you by another laboratory are externally provided (6.6). Report them as such, and check the provider's own scope before you rely on it.
- Unlisted work cannot borrow the accreditation. Where a report mixes the two, your body's rules on referring to accreditation say how to mark each result.
Sites, shared services and other interfaces
A multi-site laboratory decides which sites go on the schedule and which activities each site performs. Every listed site is assessed, and the assessors will want to see the activities that take place there, not only the head office's records of them. On-site and mobile work at the customer's premises is a location in its own right.
Look also at what sits outside the laboratory but shapes its results: a group function that calibrates your equipment, a central IT team running the laboratory information system, a purchasing department that selects reference materials. These services stay outside the scope, yet the laboratory remains responsible for them under the requirements for externally provided products and services and for control of data and information management (7.11). Assessors follow a result back through them.
How the scope moves assessment effort and cost
The accreditation body plans its assessment from the scope. Each discipline usually calls for a technical assessor with matching expertise, alongside the lead assessor. Each listed method is a candidate for witnessing, and each additional site adds travel and assessment time. A long, varied schedule means more assessors and more days, at the first assessment and at every surveillance visit.
The same logic runs through preparation. Every line on the schedule needs its own validation or verification, uncertainty evaluation and, where available, proficiency testing, and proficiency rounds follow the provider's calendar. A tightly written scope is the most effective lever a laboratory has on both the programme length and the running cost. See the cost guide and the timeline guide for the detail.
Scope mistakes that cost time
- Too wide. Applying for every method on the equipment list, so validation files are thin and the assessment finds gaps across the board.
- Too narrow. Leaving out the one method the main customer or regulator relies on, which forces an extension soon after the grant.
- Ranges beyond the data. Claiming the range printed in the standard instead of the range the laboratory has shown it can achieve.
- Silent deviations. Listing a standard method while the bench procedure differs from it, turning a verification into a validation the laboratory never did.
How Sterling helps you write the scope
A scope the evidence supports, covering the results your customers rely on.
Our expert and experienced consultants start from who uses your results, list the methods that matter to them, and check each line against the validation, uncertainty, traceability and proficiency evidence you hold. The gap analysis then shows what each line still needs, so the application goes in with a scope you can defend at the assessment. We ensure successful accreditation for our clients. The accreditation body assesses your laboratory and grants the scope; Sterling does not accredit.
Book a free discovery callISO 17025 scope of accreditation: questions buyers ask
Where can I see a laboratory's scope of accreditation?
On the accreditation body's public directory, where each accredited laboratory's schedule is published. The verification guide explains how to read it and what to check.
Should our first scope be fixed or flexible?
Fixed, in most cases. Bodies generally want to see a laboratory validate methods reliably before they allow it to add methods itself, and a fixed scope keeps the first assessment focused.
Can the range on our scope be wider than the one we validated?
No. The range should reflect the verification or validation data the laboratory holds, and assessors compare the two.
Does a revised test standard change our scope?
Usually it does. On a fixed scope, moving to the new version is normally handled as a change the body must accept; on a flexible scope, your procedure may allow it. Ask your body how it treats revised standards before the old version is withdrawn.
Can one scope cover several sites?
Yes, if the body's schedule shows which activities each site performs and every listed site has been assessed for those activities.
Does Sterling write the scope for us?
We draft it with you and test every line against your evidence; the laboratory applies, and the accreditation body decides what is granted.
More ISO/IEC 17025 guides
Every guide in the ISO/IEC 17025 series is written for one question and links back to the hub for the whole picture.
- ISO/IEC 17025 accreditation process, step by step
- ISO/IEC 17025 accreditation cost: what moves the price
- ISO/IEC 17025 documentation: what you need
- ISO/IEC 17025 internal audit: the technical clauses
- Measurement uncertainty under ISO/IEC 17025: a budget you can defend
- Method verification and validation under ISO/IEC 17025
- ISO 17025 nonconformities
- ISO 17025 proficiency testing
- ISO/IEC 17025 requirements explained, clause by clause
- ISO 17025 accreditation in Singapore and South East Asia
- ISO 17025 for small laboratories
- ISO 17025 surveillance and reassessment
Keep exploring
Explore more about ISO/IEC 17025
Watch a video on how to implement ISO 17025 here.
Watch a video on ISO 17025 requirements here.
ISO/IEC 17025 resources
Services for ISO/IEC 17025
Talk to a consultant
Write a scope of accreditation your evidence supports and your customers rely on.
Tell us your goals, locations and target date. We’ll help you define the scope, timeline and next steps.
hello@sterling-grc.com+1 201 263 4443 (global headquarters) · your nearest office and its number are at the top of this page
What would you like to achieve?
Choose a starting point. The enquiry form takes about a minute.
Book my free discovery callContinue to the enquiry form to share your details. Please do not include passwords, sensitive personal information or confidential files. A Sterling consultant replies within one working day. Privacy notice