ISO 17025 internal audit: how a laboratory audits the technical clauses, with vertical and witnessing audits

Clause by clause: ISO/IEC 17025:2017 requirements: every clause explained →

Where internal audit sits in ISO/IEC 17025:2017, what to sample under clauses 6 and 7, how to run horizontal, vertical and witnessing audits, who should audit, and the record an assessor follows.

  • Updated September 2026
  • 8-minute read
  • Part of the ISO/IEC 17025 hub

Key points

  • ISO 17025 internal audit sits in clause 8, under 8.8 Internal audits for Option A, but its real test is clauses 6 and 7: the people, equipment, traceability, methods, uncertainty and reports behind each result.
  • A laboratory needs three kinds of audit: horizontal audits of one requirement across the laboratory, vertical audits that trace one reported result back to its origins, and witnessing audits that watch the work being done.
  • Technical auditors must understand the methods they audit; a quality manager working alone rarely finds the gaps a technical assessor will.
  • Findings that show a result may have been wrong go through nonconforming work as well as corrective action, and the whole record feeds the management review.

What does ISO/IEC 17025 require of an internal audit?

ISO 17025 internal audit is the laboratory checking itself, on a planned basis, against the standard, its own procedures and the scope it holds or is applying for. Under ISO/IEC 17025:2017, the laboratory chooses how to run its management system. Option A lists the minimum elements in clause 8, and one of them is 8.8 Internal audits. Option B lets a laboratory with a management system meeting ISO 9001 rely on it, provided that system supports and demonstrates consistent fulfilment of clauses 4 to 7.

That last condition is the one most laboratories miss. An audit that looks only at document control, records and corrective action has audited the paperwork around the laboratory, not the laboratory. The requirements that decide whether a result can be trusted sit in clause 6 (resources) and clause 7 (process), and that is where the programme should spend most of its time.

ISO 19011:2026 gives general guidance on audit principles, programmes, conduct and auditor competence; the laboratory adds the technical layer itself. The full list of clauses is in the ISO/IEC 17025 requirements guide.

The technical clauses: what to audit and what to sample

A useful audit asks one question of every technical requirement: could we prove this for a result picked at random today?

Area (clause title)What the auditor checksWhat to sample
6.2 PersonnelCompetence criteria per method, training, supervision and a current authorisation for each taskThe analyst named on three recent reports and their authorisation dates
6.3 Facilities and environmental conditionsConditions the methods specify are monitored, recorded and acted on when out of limitsTemperature or humidity logs for the days the sampled work was done
6.4 Equipment and 6.5 Metrological traceabilityCalibration status, intermediate checks, labelling, and a traceability chain to the SIThe instrument used for a sampled result, its certificate and the calibration provider's accreditation
7.2 Selection, verification and validation of methodsStandard methods verified, in-house or modified methods validated, the current version in useThe verification or validation file for one method in scope
7.6 Evaluation of measurement uncertaintyA budget exists for each method where required and matches the way the work is done nowOne budget checked against the current equipment and procedure
7.7 Ensuring the validity of resultsQuality control data reviewed, proficiency testing planned, unsatisfactory results investigatedControl charts for a month and the last proficiency testing report with its follow-up
7.8 Reporting of resultsEvery required element present, statements of conformity using the agreed decision ruleFive issued reports or calibration certificates
7.11 Control of data and information managementCalculations, spreadsheets and software checked before use, access controlledOne calculation re-performed by hand

Clauses 4 and 5 need a visit too: the impartiality risk assessment should reflect the laboratory as it is now. The detail of equipment, uncertainty and methods is covered in the measurement uncertainty and method verification and validation guides.

Horizontal, vertical and witnessing audits

Horizontal audits

A horizontal audit takes one requirement and follows it across every section: equipment calibration status in all rooms, for example, or authorisations for every analyst. It spots systemic lapses quickly but never shows whether the pieces connect.

Vertical audits

A vertical audit starts from a single issued report or calibration certificate and works backwards. Take last month's certificate for a digital thermometer. The auditor confirms the decision rule was agreed at contract review, the item was identified on receipt, the procedure was the current verified version and the technician was authorised on that date. The reference standard should have been in calibration with a traceable certificate, the environmental record in limits, the raw readings should reproduce the reported values, and the stated uncertainty should match the budget. One missing link and the result cannot be defended, however good the procedures look.

Vertical audits are the type most often skipped because they take longer. Plan at least one per discipline in scope.

Witnessing audits

A witnessing audit watches an analyst or technician perform a test or calibration from start to finish, with the method open beside the auditor. It catches what records cannot: a step done from memory, a shortened warm-up, a skipped check standard. Accreditation bodies witness work during assessment, so witness yourself first, covering each authorised person over the programme.

Building the internal audit programme

The programme is the plan that decides what is audited, when, how and by whom across a defined period. Build it from the scope, not the clause list: every method group, authorised signatory, site and clause should appear in the cycle, with methods that had a complaint, a poor proficiency testing result or a new analyst coming round sooner.

Sterling's recommended pattern for a single-site laboratory preparing for first accreditation is shown below. The standard does not prescribe a frequency; the laboratory sets its own and must be able to justify it.

PeriodAudit typeCoverage
Quarter 1HorizontalClauses 4 and 5, 6.2 personnel and 6.4 equipment across all sections
Quarter 2Vertical and witnessingOne reported result per discipline traced back; witnessing of half the authorised staff
Quarter 3Horizontal7.1 contract review, 7.6 uncertainty, 7.7 validity of results, 7.11 data control
Quarter 4Vertical, witnessing and management systemRemaining disciplines and staff; clause 8 elements; follow-up of earlier findings

Before you apply, the whole programme should have run once, with findings closed. This sequence appears in the ISO/IEC 17025 implementation plan.

Who audits: competence and independence in a laboratory

Laboratory auditing needs audit technique and technical knowledge of the method, enough to know that a missing blank or an out-of-date correction factor matters. Most laboratories pair the quality manager with a technical peer from a different section, so that each audit has both.

Auditors should not audit their own work. Cross-auditing between sections works well; where one person runs a whole discipline, bring in a sister laboratory or an external auditor for those methods. The competence of each auditor should be defined and recorded like any other laboratory role, which is covered in the ISO/IEC 17025 training guide.

How Sterling supports your internal audit

A first internal audit that reaches the bench, not only the files

Our expert and experienced consultants build the programme from your scope, train your internal auditors as applicable to your standard, and conduct or facilitate the first round with them, including vertical and witnessing audits of your methods. The programme, findings and corrective actions run in Effivity, included with the engagement at no additional cost. We ensure successful accreditation for our clients.

Book a free discovery call

Findings, corrective action and the audit record

Write every finding in three parts: the requirement, the evidence seen, and the gap between them. "Uncertainty not adequate" helps no one. "The budget for the conductivity method still lists the meter replaced in March, so the reported uncertainty on reports since then is not supported" can be investigated and closed.

If the audit shows a reported result may have been affected (an unauthorised analyst, an expired reference standard, a calculation error), treat it under 7.10 Nonconforming work as well: evaluate the impact on results already issued and decide whether customers need to be told. The root cause then goes through 8.7 Corrective actions, and closure is verified against new evidence, not a statement that it was fixed.

The record is short: the programme, each audit plan, notes naming the reports, instruments and people sampled, the findings, corrective actions with evidence of closure, and the summary for management. That summary is one of the inputs to the ISO/IEC 17025 management review.

The body's assessors decide on accreditation; your auditors improve the laboratory. What happens during the body's visit is set out in The ISO/IEC 17025 assessment, and how its findings are closed in ISO/IEC 17025 nonconformities.

ISO 17025 internal audit: questions buyers ask

How often does a laboratory need to do internal audits under ISO 17025?

The laboratory sets the frequency itself and must be able to justify it against risk and scope. Most cover every clause, discipline and authorised person once a year, with extra audits after complaints or method changes.

Can our quality manager audit the whole laboratory alone?

Only for the management system clauses, in most cases. Auditing methods, uncertainty and equipment needs technical knowledge of the work, so pair the quality manager with a technical auditor from another section or an outside expert.

What is a vertical audit in an ISO 17025 laboratory?

A vertical audit picks one issued report or calibration certificate and traces it back through contract review, item handling, method, analyst authorisation, equipment, environment, raw data, uncertainty and quality control.

Do we have to audit every method before applying for accreditation?

Every method in the scope you apply for should be covered by the programme before application, at least through a vertical or witnessing audit.

We use Option B with ISO 9001. Is our ISO 9001 internal audit enough?

Not on its own. Option B relies on your ISO 9001 system only if it supports and demonstrates consistent fulfilment of clauses 4 to 7, so the audit programme must still reach the technical clauses and the methods in scope.

Can Sterling carry out our internal audits?

Yes. Our consultants can conduct or facilitate internal audits as agreed in the engagement, usually alongside your own auditors. Sterling does not accredit; the accreditation body assesses and decides.

ISO/IEC 17025 implementation

ISO/IEC 17025 requirements

Talk to a consultant

Plan an internal audit that reaches your methods, not only your procedures.

Tell us your goals, locations and target date. We’ll help you define the scope, timeline and next steps.

hello@sterling-grc.com

+1 201 263 4443 (global headquarters) · your nearest office and its number are at the top of this page

What would you like to achieve?

Choose a starting point. The enquiry form takes about a minute.

Your goal
Book my free discovery call

Continue to the enquiry form to share your details. Please do not include passwords, sensitive personal information or confidential files. A Sterling consultant replies within one working day. Privacy notice