ISO 13485 guide
ISO 13485 medical device file and design file: what goes in each and how they connect
What ISO 13485:2016 expects in the medical device file under clause 4.2.3 and the design and development file under clause 7.3.10, how the two connect at design transfer and change, a worked example, and the mistakes auditors find.
Key points
- ISO 13485:2016 requires two separate files for each device type or family: the medical device file under clause 4.2.3 and the design and development file under clause 7.3.10.
- The medical device file describes the device and how it is made, checked, stored, delivered and, where relevant, installed and serviced. The design file proves how the design came to be what it is.
- The two files meet at design transfer: approved design outputs become the specifications and procedures the medical device file points to, and every later change has to reach both.
- Neither file is a regulatory submission. Regulators and notified bodies have their own dossier formats, which draw on these files but do not replace them.
What are the medical device file and the design file?
An ISO 13485 medical device file is the controlled set of documents that defines one device type or family: what it is, what it is for, how it is labelled, what it must meet and how it is produced and checked. The design and development file is its companion: the records showing that the design was planned, specified, reviewed, verified, validated, transferred and changed under control. One answers "what are we making?", the other "how do we know the design is right?"
The standard's introduction adds that anything required to be documented must also be established, implemented and maintained, so a file correct at launch and never touched again does not meet the requirement.
| Question | Medical device file (4.2.3) | Design and development file (7.3.10) |
|---|---|---|
| What it proves | The device is defined and produced to a controlled specification | The design was developed and changed under control |
| Who keeps one | Every organisation whose scope covers the device type or family | Organisations with design and development responsibility |
| Typical content | Description, intended use, labelling, specifications, production, packaging, storage, handling, distribution, measuring and monitoring procedures, installation and servicing where relevant | Plans, inputs, outputs, review minutes, verification and validation reports, transfer records, change records |
| When it changes | Whenever a specification, label, supplier-controlled step or procedure changes | Whenever the design changes, with the evaluation of that change |
What goes in the medical device file
Clause 4.2.3 lists the content for each device type or family. In practice a good file is an index, one page long, that points to controlled documents already held in the document system, rather than a binder of copies that drift out of date.
Description, intended use and labelling
A general description of the device, its intended use or purpose, and its labelling, including instructions for use. Keep the intended use identical to the wording in the regulatory registration.
Product specifications
Drawings, material specifications, software versions, performance characteristics and acceptance criteria.
Production, distribution, measuring and monitoring
The specifications or procedures for each of these, including the steps done by suppliers, together with the measuring and monitoring procedures: inspection, test methods and release criteria. Where a contract steriliser or packaging house performs a step, the file points to the agreement and the specification that controls it.
Installation and servicing, as appropriate
Installation requirements and servicing procedures for equipment that needs them. Where neither applies, say so in one line.
What goes in the design and development file
Clause 7.3.10 closes the design and development section of ISO 13485 by requiring a file for each device type or family that brings together the records of the design work. Those records come from every step that the section controls:
- Planning: the design plan with its stages, responsibilities and planned reviews, kept current as the project moves.
- Inputs: the requirements the design must meet, including safety, performance and regulatory requirements and the outputs of risk management, reviewed and approved.
- Outputs: drawings, specifications, software, labelling and acceptance criteria, in a form that can be checked against the inputs.
- Reviews: minutes that name who attended, what was decided and the actions raised.
- Verification and validation: protocols, results and conclusions showing outputs meet inputs and the device meets user needs and intended use.
- Transfer (7.3.8): evidence that the outputs were confirmed suitable for production before routine manufacture began.
- Changes (7.3.9): each change with its review, the evaluation of its effect and its verification or validation.
The risk management file, run to the process clause 7.1 requires and usually built with ISO 14971, sits alongside. It is referenced from the design file, not copied into it. Software adds IEC 62304 life-cycle records, covered in ISO 13485 for medical device software.
An organisation with no design responsibility, such as a distributor or a contract manufacturer building to a customer's drawings, keeps no design file of its own. Its scope should state clearly where design sits.
How the two files connect
The link is design transfer. When the design is released, its approved outputs become the specifications, labels and procedures the medical device file indexes. After that, any change, whether from production, a complaint or a supplier, is evaluated in the design file, reflected in the medical device file and checked against the risk management file.
| Event | Design file records | Medical device file updates |
|---|---|---|
| Design released for production | Transfer review, final verification and validation reports | Specifications, labels and procedures issued at their approved revision |
| Supplier changes a material | Change evaluation, risk review, any repeat verification | Material specification and approved supplier reference |
| Complaint shows a labelling weakness | Change record linked to the complaint and CAPA | Revised instructions for use |
Auditors test this link both ways: from a specification back to an approved design output, and from an output forward to the document production uses.
A worked example: a sterile single-use extension set
Take a small manufacturer that designs a sterile single-use infusion extension set in three lengths, moulds the connectors in house and sends finished packs to a contract steriliser.
The three lengths share materials, design, process and intended use, so they form one family with one medical device file. The index lists the description and intended use; the label and instructions for use; the drawing set and material specifications; the moulding, assembly and packaging procedures; the storage and distribution instruction; the agreement and specification for the sterilisation step; and the inspection and release procedures. Installation and servicing are marked as not applicable, with the reason.
The design file holds the plan, the input list traced to each output, the review minutes from concept to release, the verification reports for connection strength, flow and leakage, the usability validation, the transfer review and the change log. The risk management file is referenced, its hazards traced to the inputs that control them.
Batch and sterilisation records are not in either file. They are records of production under clause 7.5.5, kept per batch and traced to it; the medical device file points to the procedures that generate them.
The files and regulatory submissions
Neither file is the dossier a regulator reviews. Class B, C and D devices registered with the Philippine FDA use the ASEAN Common Submission Dossier Template; CE marking under the EU MDR needs a notified body for most device classes, with its own documentation expectations; and the FDA's QMSR incorporates ISO 13485:2016 but states that compliance with ISO 13485 alone does not fully satisfy it. A well-kept pair of files is the source those dossiers are assembled from. The US picture is covered in ISO 13485 and the FDA QMSR: what to check.
Common mistakes and what auditors test
Certification body auditors usually sample one or two families and follow them end to end. Most findings on these files come from a few habits:
- The design history lives in engineers' e-mail and test folders and is assembled a week before the audit.
- Risk management is a spreadsheet completed after the design was frozen, so no design input can be traced to a hazard.
- One file covers devices that do not share a design or process, and the grouping cannot be justified.
- Supplier-performed steps, above all sterilisation and packaging, have no controlling specification in the file.
- Changes are made on the line and never reach the design file.
How Sterling builds the files
Files built from your engineering records, kept live in Effivity
Sterling's expert and experienced consultants work with your engineers to group devices into defensible families, index each medical device file, and rebuild the design file from the records you already hold. Both files live in Effivity, included with the engagement at no additional cost, so a change updates them together. We ensure successful certification for our clients; an accredited certification body audits and decides.
Book a free discovery callMedical device file and design file: questions buyers ask
Is the design and development file the same as a design history file?
It plays the same role: many engineering teams call it the design history file, while ISO 13485:2016 names it the design and development file under clause 7.3.10. What matters to the auditor is the content, the records of planning, inputs, outputs, reviews, verification, validation, transfer and changes, not the label on the folder.
Can one medical device file cover several devices?
Yes, one file can cover a device family, provided the devices genuinely share their design, materials, manufacturing process and intended use. Record the reason for the grouping and split the file where variants need different procedures.
Does a contract manufacturer need a medical device file?
A contract manufacturer inside ISO 13485 scope needs a file for the devices it makes, covering the specifications and procedures it controls, even when the brand owner holds the design. It keeps a design file only if it carries design responsibility.
Where does the risk management file fit?
The risk management file is a separate record required by clause 7.1, referenced from the design file and read again at every change. Its hazards should trace to design inputs and its controls to verified outputs.
How long does it take to build the files for an existing device?
It depends on how much of the design history exists: where test reports and review records are on hand, indexing and closing gaps takes weeks; where they are missing, verification may have to be repeated. The certification process guide places this work in the overall plan.
More ISO 13485 guides
Every guide in the ISO 13485 series is written for one question and links back to the hub for the whole picture.
- ISO 13485 case study
- ISO 13485 certification process, step by step
- ISO 13485 certification cost: what moves the price
- The medical device file and the design file: what goes in them
- Effivity for ISO 13485: the guide
- ISO 13485 internal audit: planning, conduct and reporting
- ISO 13485 and the FDA QMSR: what to check
- ISO 13485 vs ISO 9001: what the device standard adds
- ISO 13485 requirements explained, clause by clause
- ISO 13485 training: who needs what
Keep exploring
Explore more about ISO 13485
Watch a video on how to implement ISO 13485 here.
Watch a video on ISO 13485 requirements here.
ISO 13485 resources
Services for ISO 13485
Industries
Talk to a consultant
Build device and design files your engineers keep and your auditors can follow.
Tell us your goals, locations and target date. We’ll help you define the scope, timeline and next steps.
hello@sterling-grc.com+1 201 263 4443 (global headquarters) · your nearest office and its number are at the top of this page
What would you like to achieve?
Choose a starting point. The enquiry form takes about a minute.
Book my free discovery callContinue to the enquiry form to share your details. Please do not include passwords, sensitive personal information or confidential files. A Sterling consultant replies within one working day. Privacy notice